Remote Control
R&S
®
FPC
240
User Manual 1178.4130.02 ─ 07
Parameters:
<State>
ON | 1
Turns on the third order intermodulation measurement.
OFF | 0
Performs a basic spectrum measurement.
*RST:
OFF
Example:
//Measure the TOI
CALC:MARK:FUNC:TOI ON
Manual operation:
See
19.12.1.4
Occupied Bandwidth
The following remote commands are exclusive to the occupied bandwidth measure-
ment.
CALCulate<n>:MARKer<m>:FUNCtion:OBW:BANDwidth
.................................................. 240
CALCulate<n>:MARKer<m>:FUNCtion:OBAN:BANDwidth
.................................................240
CALCulate<n>:MARKer<m>:FUNCtion:OBW:BANDwidth:PCT
...........................................240
CALCulate<n>:MARKer<m>:FUNCtion:OBAN:BANDwidth:PCT
......................................... 240
CALCulate<n>:MARKer<m>:FUNCtion:OBW:BANDwidth
<ChannelBW>
CALCulate<n>:MARKer<m>:FUNCtion:OBAN:BANDwidth
<ChannelBW>
This command defines the channel bandwidth for occupied bandwidth measurements.
Prerequisites for this command
●
Turn on occupied bandwidth measurement (
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<ChannelBW>
<numeric value>
*RST:
3.84 MHz
Default unit: Hz
Example:
//Define a channel bandwidth
CALC:MARK:FUNC:OBW:BAND 1 MHZ
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:OBW:BANDwidth:PCT
<Percentage>
CALCulate<n>:MARKer<m>:FUNCtion:OBAN:BANDwidth:PCT
<Percentage>
This command defines the percentage of the total power that defines the occupied
bandwidth.
Suffix:
<n>, <m>
.
irrelevant
Remote Commands of the Spectrum Application