The Spectrum Application
R&S
®
FPC
72
User Manual 1178.4130.02 ─ 07
The displayed result shows the highest channel power that has been measured
(indicated by the "Max Power" result in the diagram header.
4. You can return to the normal power display by selecting the "Clear / Write" menu
item instead.
Measuring the channel power density
By default, the R&S
FPC the channel power result corresponds to the complete chan-
nel bandwidth you have selected. You can also display the channel power that corre-
sponds to a bandwidth of 1
Hz. In that case, the results are no longer displayed in
dBm, but in dBm/Hz.
Using this evaluation method, you can measure the signal / noise power density, for
example.
1. Press the "Meas" key to open the measurement menu.
2. Select the "Power Display" menu item.
3. Select the "Channel Power / Hz" menu item.
The displayed result shows the channel power measured in a bandwidth of 1
Hz
(indicated by the different unit in the channel power result in the diagram header).
4. You can return to the normal power display by selecting the "Channel Power / Hz"
menu item again.
Selecting the unit for power display
The R&S
FPC can apply different units to the power output.
1. Press the "Meas" key to open the measurement menu.
2. Select the "Power Unit" menu item to open the power unit menu.
3. Select the unit you want.
The R&S
FPC displays the signal level in the selected unit.
Note:
The default unit is dBm.
Selecting a standard
You can store the main measurement settings as a standard file. When you load such
a standard file, the R&S
FPC automatically configures the measurement regarding
channel and general measurement settings.
Standard files for common telecommunications standards are already provided with
the R&S
FPC firmware.
If you want to measure a different standard, you can also create custom standard defi-
nition files with the R&S
InstrumentView software and transfer them to the R&S
FPC.
For more information, refer to the documentation of the R&S
InstrumentView.
1. Press the "Meas" key to open the measurement menu.
2. Select the "Standard" menu item.
Measurements and Result Displays