User Manual 1178.4130.02 ─ 07
DUT
R&S FPC
Port 1
(RF Out)
Figure 18-2: Setup of a reflection S11 measurement
If you measure the reflection on a system with two or more system components, the
results are an aggregate over these components. Therefore you can only see the
aggregated magnitude of the reflected power in the selected frequency range.
Measuring the reflection
1. Connect the DUT to the R&S
FPC as shown in
2. Define the frequency parameters for the measurement.
For more information, see
Chapter 18.2, "Frequency Configuration"
3. Press the "Meas" key to enter the measurement menu.
4. Select the "Reflection S11" menu item to start the measurement.
5. Calibrate the reflection measurement.
For more information, see
6. After calibration, start the actual measurement (single sweep or continuous
sweep).
18.1.3
1-Port Cable Loss Measurement
Access
: "Meas" > "1-Port Cable Loss"
The cable loss measurement is a reflection measurement on an open cable. It evalu-
ates the amount of RF energy that a signal loses while traveling through a cable. The
amount of lost energy depends on the frequency and the length of the cable.
Cable
R&S FPC
Port 1
(RF Out)
Open
Figure 18-3: Setup of a 1-port cable loss measurement
By default, the R&S
FPC evaluates the cable loss over its entire frequency range (full
span of the R&S
FPC). When you reduce the span, make sure to define a span that is
appropriate for the measurement, especially for high center frequencies. Also make
sure to consider the cable length and the wavelength of the selected center frequency.
For a typical cable loss trace, select a span that is large enough to evaluate at least
one peak and one minimum. Otherwise, the result might not be valid.