![Rohde & Schwarz FPC Series User Manual Download Page 339](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824339.webp)
Remote Control
R&S
®
FPC
336
User Manual 1178.4130.02 ─ 07
Parameters:
<State>
ON | 1
Frequency settings are coupled. When you change the fre-
quency or span in the DTF measurement, they also change in
the other VNA measurements (reflection, transmission and cable
loss).
OFF | 0
Frequency settings of the DTF measurement are independent.
When you change the frequency or span in the DTF measure-
ment, they do not change in the other VNA measurements
(reflection, transmission and cable loss).
*RST:
ON
Example:
//Decouple frequency and span
FREQ:SETT:COUP:ENAB OFF
Manual operation:
See
[SENSe:]FREQuency:SPAN:AUTO
<State>
This command turns the automatic calculation of the ideal span on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF | 0
Example:
//Turn on automatic span
FREQ:SPAN:AUTO ON
Manual operation:
See
See
19.16.4
Bandwidth
Commands to configure the bandwidth described elsewhere.
●
Resolution bandwidth:
[SENSe:]BANDwidth[:RESolution]
[SENSe:]BANDwidth[:RESolution]:AUTO
19.16.5
Sweep
Commands to configure the sweep described elsewhere.
●
Sweep mode:
●
Initiate measurement:
.................................................................................................337
Remote Commands of the VNA Application