Remote Control
R&S
®
FPC
259
User Manual 1178.4130.02 ─ 07
Parameters:
<Unit>
DBM | DBUV | DBMV | V | W | DUVM | DUAM | V_M | W_M2
Note that the availability of units depends on the operating
mode.
*RST:
dBm
Example:
//Select displayed unit
UNIT:POW DBUV
Manual operation:
See
See
19.12.4
Bandwidth
.........................................................................................259
[SENSe:]BANDwidth[:RESolution]
................................................................................... 259
[SENSe:]BWIDth[:RESolution]:AUTO
............................................................................... 260
[SENSe:]BANDwidth[:RESolution]:AUTO
..........................................................................260
............................................................................................ 260
........................................................................................ 260
...................................................................................260
[SENSe:]BWIDth[:RESolution]
<Bandwidth>
[SENSe:]BANDwidth[:RESolution]
<Bandwidth>
This command selects the resolution bandwidth (or measurement bandwidth).
Effects of this command
●
Turns off automatic selection of resolution bandwidth (
Parameters:
<Bandwidth>
<numeric value> (integer)
Value adjustable in 1 - 3 steps (1
Hz, 3
Hz, 10
Hz, etc.).
In addition, you can select a 200 kHz bandwidth.
Range:
1 Hz to 3 MHz
*RST:
- (AUTO is set to ON)
Default unit: Hz
Example:
//Define resolution bandwidth
BAND 100KHZ
Manual operation:
See
See
See
Remote Commands of the Spectrum Application