19.12.1.2
Channel Power
The following remote commands are exclusive to the channel power measurement.
CALCulate<n>:MARKer<m>:FUNCtion:CPOWer:BANDwidth
............................................. 237
CALCulate<n>:MARKer<m>:FUNCtion:CPOWer:MODE
....................................................237
CALCulate<n>:MARKer<m>:FUNCtion:CPOWer:UNIT
...................................................... 238
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult:PHZ
.............................................. 238
CALCulate<n>:MARKer<m>:FUNCtion:CPOWer:BANDwidth
<Bandwidth>
This command defines the channel bandwidth for channel power measurements.
Prerequisites for this command
●
Select the channel power measurement (
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<Bandwidth>
<numeric value>
*RST:
3.84 MHz
Default unit: Hz
Example:
//Define the channel bandwidth
CALC:MARK:FUNC:CPOW:BAND 10MHZ
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:CPOWer:MODE
<DisplayMode>
This command selects the display mode for channel power measurements.
Prerequisites for this command
●
Select the channel power measurement (
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<DisplayMode>
CLR
Display the currently measured value ("Clear / Write").
MAX
Display the highest measured value ("Max Hold").
*RST:
CLR
Example:
//Select clear/write trace mode
CALC:MARK:FUNC:CPOW:MODE CLR
Manual operation:
See