Remote Control
R&S
®
FPC
339
User Manual 1178.4130.02 ─ 07
NADMittance
standardized admittance in complex format (real +
imaginary)
NIMPedance
standardized impedance in complex format (real +
imaginary)
NORMal
normal marker
RPDB
reflection coefficient in complex format (magnitude
(dB) + phase)
RPL
reflection coefficient in complex format (magnitude
(lin) + phase
RSCalar
reflection coefficient in complex format (real + imagi-
nary)
CALCulate<n>:MARKer<m>:Y?
This command queries the measurement results at the marker position.
If necessary, the corresponding marker is activated first.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
Suffix:
<n>
.
irrelevant
<m>
Return values:
<Position>
<numeric value>
The unit of the return value depends on the measurement mode
and format you have selected.
In the Smith chart, the return value also depends on the marker
format you have selected with
Example:
//Turn on single sweep mode and marker 2
INIT:CONT OFF
CALC:MARK2 ON
//Start a sweep and wait for the end
INIT;*WAI
//Query the position of marker 2
CALC:MARK2:X?
CALC:MARK2:Y?
Usage:
Query only
Manual operation:
See
Remote Commands of the VNA Application