Remote Control
R&S
®
FPC
202
User Manual 1178.4130.02 ─ 07
*OPC with a service request
1. Set the OPC mask bit (bit no. 0) in the ESE: *ESE 1
2. Set bit no. 5 in the SRE:
*SRE 32
to enable ESB service request.
3. Send the overlapped command with
*OPC
.
4. Wait for a service request.
The service request indicates that the overlapped command has finished.
*OPC? with a service request
1. Set bit no. 4 in the SRE:
*SRE 16
to enable MAV service request.
2. Send the overlapped command with
*OPC?
.
3. Wait for a service request.
The service request indicates that the overlapped command has finished.
Event status register (ESE)
1. Set the OPC mask bit (bit no. 0) in the ESE:
*ESE 1
2. Send the overlapped command without
*OPC
,
*OPC?
or
*WAI
.
3. Poll the operation complete state periodically (with a timer) using the sequence:
*OPC; *ESR?
A return value (LSB) of 1 indicates that the overlapped command has finished.
19.1.6
Suffix Overview
Table 19-5: Overview of SCPI suffixes
Suffix
Description
Range
<m>
Selects a marker
1...6
<n>
Selects a window
irrelevant
<t>
Selects a trace
1...2
<td>
Selects the transducer type (pri-
mary or secondary)
1...2
19.2
Common Commands
Common Commands