Remote Control
R&S
®
FPC
230
User Manual 1178.4130.02 ─ 07
Example:
//Query calibration type
CAL:MODE?
Usage:
Query only
Manual operation:
See
CALibration<n>:STARt?
<CalibrationMethod>
This command selects a calibration method and initiates calibration.
After the command has been sent, it returns instructions on what to do to continue the
calibration process and finish it successfully.
Suffix:
<n>
.
irrelevant
Parameters for setting and query:
<CalibrationMethod>
S11Cal
Full 1-port calibration.
Unavailable for S21 transmission measurements.
S11easy
Easy 1-port calibration.
Unavailable for S21 transmission measurements.
S21Norm
Normalization for S21.
Available for S21 transmission measurements.
Return values:
<Instructions>
String containing the instructions for the next calibration step.
You can find a list possible instructions in the table below.
Example:
//Start full 1-port calibration
MEAS:MODE REFL
CAL:STAR? S11C
//would return the instruction
'Connect Open to Port 1'
Usage:
Query only
Manual operation:
See
See
Table 19-7: Return values: Possible instructions for calibration
'Connect Open to Port 1'
Connect the open to port 1 before resuming calibration.
'Connect Short to Port 1'
Connect the short to port 1 before resuming calibration.
'Connect Load to Port 1'
Connect the load to port 1 before resuming calibration.
'Connect Through'
Connect a through to ports 1 and 2.
'Disconnect DUT from port 1'
'Calibration Done'
Calibration has been finished. No further actions required.
Calibration