The Spectrum Application
R&S
®
FPC
83
User Manual 1178.4130.02 ─ 07
The R&S
FPC opens an input field to define the channel bandwidth.
3. Enter the channel bandwidth you need.
Note:
The minimum channel bandwidth that you can set is 2
kHz at a span of
1
kHz.
Selecting a standard
You can store the main measurement settings as a standard file. When you load such
a standard file, the R&S
FPC automatically configures the measurement regarding
channel and general measurement settings.
Standard files for common telecommunications standards are already provided with
the R&S
FPC firmware.
If you want to measure a different standard, you can also create custom standard defi-
nition files with the R&S
InstrumentView software and transfer them to the R&S
FPC.
For more information, refer to the documentation of the R&S
InstrumentView.
1. Press the "Meas" key to open the measurement menu.
2. Select the "Standard" menu item.
The R&S
FPC opens a dialog box to select the standard.
3. Select one of the available standards.
The R&S
FPC loads the configuration of the selected standard. It automatically
sets the span, resolution bandwidth, video bandwidth, sweep time and detector
required by the standard.
If the measurement is not according to the selected standard anymore (for exam-
ple if you change a parameter), the R&S
FPC shows a blue dot in front of the stan-
dard display.
Defining the reference level
is the power level that the R&S
FPC expects at the RF input. When
you select the reference level, make sure that you do not overload the R&S
FPC by
applying a signal whose power exceeds the maximum reference level.
As the power is measured with a small
bandwidth, it is still possible to overload the R&S
FPC, even if the trace is within the
measurement diagram. To prevent an overload, measure the occupied bandwidth with
the largest resolution bandwidth possible using the peak detector. If you set these
parameters, it is not possible for the trace to exceed the reference level.
To simplify operation and to prevent incorrect measurements, the R&S
FPC has an
automatic routine for setting the reference level.
1. Press the "Meas" key to open the measurement menu.
2. Select the "Level Adjust" menu item.
The R&S
FPC performs a measurement to determine the optimal reference level.
Measurements and Result Displays