Remote Control
R&S
®
FPC
268
User Manual 1178.4130.02 ─ 07
If you are using the average trace mode, you can set the number of measurements
with
. Note that synchronization to the end of the average
count is possible only in single sweep mode.
Suffix:
<n>
.
irrelevant
<t>
Parameters:
<TraceMode>
You can turn off the trace with
For more information about the trace modes, see
AVERage
"Average" trace mode.
BLANk
"Blank" trace mode.
MAXHold
"Max Hold" trace mode.
MINHold
"Min Hold" trace mode.
VIEW
"View" trace mode.
WRITe
"Clear / Write" trace mode.
*RST:
WRITe
Example:
//Select trace mode
DISP:TRAC:MODE AVER
Manual operation:
See
See
See
DISPlay<n>[:WINDow]:TRACe<t>[:STATe]
<State>
This command turns a trace on and off.
Suffix:
<n>
.
irrelevant
<t>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF | 0
Example:
//Turn on trace 2
DISP:TRAC2 ON
Remote Commands of the Spectrum Application