![Rohde & Schwarz FPC Series User Manual Download Page 227](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824227.webp)
Remote Control
R&S
®
FPC
224
User Manual 1178.4130.02 ─ 07
MMEMory:NAME
<FileName>
This command defines the path and file name of a screenshot.
You can take the actual screenshot with
.
Parameters:
<FileName>
String containing the path and name of the file, including the file-
name extension (
.png
or
.jpg
).
The path and file name comply with Windows conventions. Note
that the folder and filename references in the string are case-
sensitive.
Example:
//Take a screenshot
MMEM:NAME 'Public\Screeshots\Test.png'
HCOP
Manual operation:
See
MMEMory:RDIRectory
<Directory>
This command deletes the indicated directory.
Setting parameters:
<Directory>
String containing the path and name of the directory you want to
delete.
The path and folder name comply with Windows conventions.
Note that the folder references in the string are case-sensitive.
Example:
//Delete a directory
MMEM:RDIR '\Public\funny'
Usage:
Setting only
Manual operation:
See
MMEMory:STORe:STATe
<1>, <TargetFile>
This command saves the current measurement configuration and results in a dataset.
Parameters:
<TargetFile>
String containing the path and name of the destination file,
including the filename extension (
.set
).
The path and file name comply with Windows conventions. Note
that the folder and filename references in the string are case-
sensitive.
Setting parameters:
1
<numeric value>
Example:
//Save the current measurement configuration
MMEM:STOR:STAT 1,'\Public\result.set'
Usage:
Setting only