Digital Demodulation
R&S
®
FPC
168
User Manual 1178.4130.02 ─ 07
2. Select the "Deviation Per Division" menu item.
3. Enter the deviation you like.
The R&S
FPC adjusts the scale of the y-axis accordingly. Note that the trace is
also compressed or enlarged accordingly.
17.5
Sweep Configuration
Access
: "Sweep"
Remote commands to control the measurement:
●
Chapter 19.14.5, "Sweep Control"
The sweep menu provides settings to control the measurement and the way data is
captured.
Selecting the measurement mode
The measurement mode defines how often the data is demodulated.
In continuous mode, the R&S
FPC demodulates the signal over and over until you stop
it deliberately. The measurement results are updated accordingly.
In single mode, the R&S
FPC demodulates the signal once and then stops without
updating the results any longer. Another measurement starts only when you deliber-
ately initiate it.
1. Press the "Sweep" key to open the sweep menu.
2. Select the "Continuous Sweep" menu item.
The R&S
FPC demodulates the signal continuously.
You can interrupt a continuous measurement by selecting the "Continuous Sweep"
menu item again.
3. Select the "Single Sweep" menu item.
The R&S
FPC demodulates the signal once and displays the results of that mea-
surement.
17.5.1
Function Reference
........................................................................................... 168
Continuous Measurement
See
"Selecting the measurement mode"
Remote command:
Measurement type:
New measurement sequence:
Sweep Configuration