Interfaces/Annex A
CTS55
1094.0006.55
A.4
E-3
Cable for local controller coupling in the case of hardware handshake
The connection of the CTS to a controller is effected with a so-called zero modem cable. In the case of
this cable, the data, control and report lines must be crossed. The following wiring diagram applies to a
controller with 9-pin or 25-pin configuration.
CTS
Controller
CTS
Controller
9-pin 9-pin
9-pin 25-pin
1
2
3
4
5
6
7
8
9
------------RxD / TxD---------------
------------TxD / RxD---------------
------------DTR /DSR--------------
------------GND / GND-------------
------------DSR / DTR--------------
------------RTS / CTS--------------
------------CTS / RTS--------------
1
3
2
6
5
4
8
7
9
1
2
3
4
5
6
7
8
9
------------RxD / TxD---------------
------------TxD / RxD---------------
------------DTR /DSR--------------
------------GND / GND-------------
------------DSR / DTR--------------
------------RTS / CTS--------------
------------CTS / RTS--------------
8
2
3
6
7
20
5
4
22
Fig. A3-3
Wiring of the data, control and report lines for hardware handshake
Summary of Contents for 1094.0006.50
Page 4: ......
Page 12: ......
Page 22: ......
Page 24: ......
Page 26: ......
Page 28: ......
Page 30: ......
Page 41: ...CTS55 Getting Started With Autotest 1094 0006 55 3 3 E 7...
Page 66: ......
Page 81: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 7 E 6...
Page 83: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 9 E 6...
Page 106: ......
Page 108: ......
Page 196: ......
Page 246: ......
Page 250: ......
Page 254: ......
Page 256: ......
Page 343: ...CTS K6 Windows Application CTSgo 1079 2001 01 87 E 3 Fig CTSgo help window...