List of Error Messages/Annex B
CTS55
1094.0006.55
B.4
E-3
Device Specific Error; sets bit 3 in the ESR register
Error code
Error test in the case of queue poll
Error explanation
-300
Device-specific error
SM3-specific error not defined in greater detail.
-350
Queue overflow
This error code is entered in the queue instead of the actual error code if the queue is full. It indicates
that an error has occurred but not been accepted. The queue can accept 5 entries.
Query Error - Error in data request; sets bit 2 in the ESR register
Error code
Error text in the case of queue poll
Error explanation
-400
Query error
General error occurring when data are requested by a query.
-410
Query INTERRUPTED
The query has been interrupted.
Example: After a query, the instrument receives new data before the response has been sent completely.
-420
Query UNTERMINATED
The query is incomplete.
Example: The instrument is addressed as a talker and receives incomplete data.
-430
Query DEADLOCKED
The query cannot be processed.
Example: The input and output buffers are full, the instrument cannot continue operation.
-440
Query UNTERMINATED after indefinite response
A query is in the same command line after a query which requests an indefinite response.
Summary of Contents for 1094.0006.50
Page 4: ......
Page 12: ......
Page 22: ......
Page 24: ......
Page 26: ......
Page 28: ......
Page 30: ......
Page 41: ...CTS55 Getting Started With Autotest 1094 0006 55 3 3 E 7...
Page 66: ......
Page 81: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 7 E 6...
Page 83: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 9 E 6...
Page 106: ......
Page 108: ......
Page 196: ......
Page 246: ......
Page 250: ......
Page 254: ......
Page 256: ......
Page 343: ...CTS K6 Windows Application CTSgo 1079 2001 01 87 E 3 Fig CTSgo help window...