Test Sequence
CTS55
1094.0006.55 4.2
E-6
In the description of the automatic sequence below it is assumed that the measurements are carried out
on two channels and that the wide tolerances are used.
In the description of the steps of the Autotest the possible causes for an error are listed. This list cannot
be regarded as complete and merely contains possible faults that are easy to eliminate e.g. flat
batteries.
Before the description of the tests here are some errors common for all test steps:
•
The mobile is switched off during a running test sequence.
In this case the error message "Detach not expected" may appear on the screen.
•
The connection is released from the mobile in a state which the release is not expected.
In this case the error message "Mobile has been released during power measurement, sensitivity
measurement, ..." appears on the screen.
•
There are problems with the signalling of the mobile (messages of the mobile are wrong).
In this case the error message "Signalling error during Location Update, Call To, ..." appears on the
screen.
Summary of Contents for 1094.0006.50
Page 4: ......
Page 12: ......
Page 22: ......
Page 24: ......
Page 26: ......
Page 28: ......
Page 30: ......
Page 41: ...CTS55 Getting Started With Autotest 1094 0006 55 3 3 E 7...
Page 66: ......
Page 81: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 7 E 6...
Page 83: ...CTS55 Getting Started With Manual Test 1094 0006 55 6 9 E 6...
Page 106: ......
Page 108: ......
Page 196: ......
Page 246: ......
Page 250: ......
Page 254: ......
Page 256: ......
Page 343: ...CTS K6 Windows Application CTSgo 1079 2001 01 87 E 3 Fig CTSgo help window...