13. Reconnect the wires to the sensors on the green luer lock side. The color of the sensor
wire is irrelevant when connecting to the sensor socket. Either color of sensor wire can
connect to either sensor socket.
Figure 7-11: Water Level Sensor Wire Sockets
Figure 7-12: Water Level Sensor Wires Inside the Instrument
14. Reconnect the CO
2
/air tubing to the luer locks on both sides of the water reservoir.
15. Replace the top access panel surrounding the top door and close the upper hinged side
door. See
.
16. Replace the lock screw in the top access panel. See
17. If your instrument includes the Transmitted Light option, move the Transmitted Light tower
back into position and replace the hinge release knob. See
18. Restart the system. See
Starting the System on page 27
19. In the MetaXpress software, do one of the following to close the instrument top door:
From the simplified menu, click Screening > Acquisition Setup > Load Plate.
From the standard menu, click Screening > Plate Acquisition Setup > Load Plate.
Summary of Contents for ImageXpress Confocal HT.ai
Page 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Page 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Page 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Page 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Page 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Page 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Page 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...