Chapter 7: Expansion Solution Options
5084773 A
167
Setting Up Koehler Illumination
To ensure high-quality illumination with brightfield or phase contrast imaging, set up the
Transmitted Light option for Koehler illumination. Do this alignment anytime there is concern
about the quality of the transmitted light images or when changing plate type or slide type.
Since the objectives are parfocal, repeating this procedure after switching objectives is
unnecessary.
Figure 7-43: Transmitted Light Overview
Item
Description
1
Field Aperture Shutter (O = open; C = closed)
2
Field Aperture Diaphragm Control
3
Condenser Focus Control
4
Condenser Adjustment Screws (Aperture Lateral Adjustment Control)
5
Condenser Turret
6
Field Stop Slider (Condenser Aperture Diaphragm Control)
Table 7-9: Transmitted Light Overview Description
Note: For the ImageXpress Confocal HT.ai system, make sure that the instrument is in
widefield mode before setting up Koehler illumination.
Summary of Contents for ImageXpress Confocal HT.ai
Page 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Page 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Page 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Page 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Page 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Page 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Page 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...