Chapter 7: Expansion Solution Options
5084773 A
171
2. Rotate the condenser turret so that the A position is selected (at the front).
Figure 7-46: Condenser Turret with Position A at the Front
3. If Koehler illumination has not been set up, turn the adjustment ring, which is located below
the arm of the Transmitted Light tower, to the middle of the aperture range.
Figure 7-47: Aperture Stop Position for Phase Ring Alignment
Item
Description
1
Label indicating aperture range
2
Field aperture diaphragm control
Table 7-10: Aperture Stop Position for Phase Ring Alignment Description