ImageXpress Confocal HT.ai High-Content Imaging System User Guide
44
5084773 A
Verifying Calibration Settings
Use the Calibrate Distances dialog to verify calibration distances.
To verify the calibration settings:
1. In the MetaXpress software, do one of the following to display the Calibrate Distances
dialog:
From the simplified menu, click Measure > Distances > Calibrate Distances.
From the standard menu, click Measure > Calibrate Distances.
2. Confirm that there are calibration settings in the dialog that match the objective settings
from the Configure Magnification dialog.
Figure 3-13: Calibrate Distances Dialog