Chapter 7: Expansion Solution Options
5084773 A
141
6. Locate the water reservoir through the top opening.
Figure 7-10: Environment Control Water Reservoir
Item
Description
1
Luer Locks
2
Water Sensors Wires
3
Water Sensor
Table 7-2: Environment Control Water Reservoir Description
7. Disconnect the CO
2
/air tubing from the luer locks on both side of the water reservoir.
8. Unplug the sensors from the green luer lock side of the water reservoir and remove the
water reservoir from the instrument.
9. Use a Phillips screwdriver to loosen the four screws securing the reservoir lid.
10. Remove the lid.
11. Fill the reservoir with up to 250 mL (8.5 oz) of deionized water to the Max Fill Line.
12. Replace the lid, tighten the screws, and then place the reservoir back in the instrument.