![Molecular Devices ImageXpress Confocal HT.ai User Manual Download Page 173](http://html1.mh-extra.com/html/molecular-devices/imagexpress-confocal-ht-ai/imagexpress-confocal-ht-ai_user-manual_1823736173.webp)
Chapter 7: Expansion Solution Options
5084773 A
173
12. Adjust the position of the Z-stage to obtain a reasonably sharp image of the objective
phase ring.
Figure 7-50: In Focus Image of Phase Contrast Ring
13. Use the region tools in the software to mark the position of the phase ring. Use the
rectangular region tool to make sure that you capture both horizontal and vertical axes.
Figure 7-51: Marking the Position of the Phase Ring
14. Do one the following to save the marked positions:
From the simplified menu, click Measure > Regions > Save Region.
From the standard menu, click Regions > Save Regions.
15. Manually rotate the condenser turret to the annulus for the phase objective (PhL for 4x, Ph1
for 10x and 20x, Ph2 for 40x).
The annulus is at the front of the condenser turret.
Phase contrast objectives typically have their specifications indicated in green letters on
the outer barrel. These specifications also include the matching annulus designation (for
example, PhL, Ph1, Ph2).
Summary of Contents for ImageXpress Confocal HT.ai
Page 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Page 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Page 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Page 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Page 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Page 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Page 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...