ImageXpress Confocal HT.ai High-Content Imaging System User Guide
146
5084773 A
Setting Up the CO
2
Tank
A pre-mixed CO
2
tank with a regulator must be set up and connected to the CO
2
inlet on the
ImageXpress Systems Power and Options Controller, which controls the flow rate of CO
2
delivered to the water reservoir within the environmental enclosure.
Before you begin, confirm that the water reservoir is set up. See
.
To set up the CO
2
:
1. Connect the regulator to the CO
2
tank.
2. Connect the tubing from the pre-mixed CO
2
tank regulator to the ImageXpress Systems
Power and Options Controller.
3. Verify that the tubing from the ImageXpress Systems Power and Options Controller to the
instrument is connected.
Figure 7-17: CO
2
Tube Connection on the Rear of the Instrument
4. Turn on the CO
2
regulator to approximately 1.03 to 1.38 bar (15 to 20 psi).
Summary of Contents for ImageXpress Confocal HT.ai
Page 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Page 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Page 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Page 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Page 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Page 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Page 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...