23. In the Focus dialog, on the Auto Focus tab, click Preview Pass to display a Preview Pass
image window that displays a graph of focus intensities compared to Z-position.
Figure 8-33: Preview Pass Image Window
The top of the red peak represents the brightest pixel of the preview pass. The top of the
green peak represents the highest focus score. If there is no peak in the graph:
In the Focus dialog, increase the Start Position value by 1000 and click Preview Pass.
Repeat until you cover a range between 6000 and 10000.
If there still is no peak in the graph, move to a new area of the plate and click Preview
Pass. If no peak after you move twice and cover a z-range from 6000 to 10000,
contact Technical Support.
24. In the Preview Pass image window, click the trace line at the top of the green line of the
first peak. The graph X, Y coordinates display as a tool tip. The X coordinate is the value
for the Z-position, 8300 in the example.
Figure 8-34: Z-Position of Peak
Summary of Contents for ImageXpress Confocal HT.ai
Page 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Page 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Page 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Page 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Page 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Page 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Page 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...