D-2
Measurement
Alarm messages
L
Cause and solution
Missed Beats
M*
Irr. Rhythm
M*
Vent. Rhythm
M*
Multif. PVC
M*
Nonsus. Vtac
M*
Pause
L*
PNP
M*
The pacer appears abnormal. Check the pacer.
PNC
M*
Resp
Resp Apnea
H
The respiration signal was so weak that the monitor cannot perform
respiration analysis. Check the patient’s condition and the Resp
connections.
Resp Artifact
H
The patient’s heartbeat has interfered with his respiration. Check the
patient’s condition and the Resp connections.
SpO
2
SpO
2
Desat
H
The SpO2 value has fallen below the desaturation alarm limit. Check
the patient’s condition and check if the alarm limit settings are
correct.
No Pulse
H
The pulse signal was so weak that the monitor cannot perform pulse
analysis. Check the patient’s condition, SpO
2
sensor and measurement
site.
CO
2
CO
2
Apnea
H
The patient stops breathing, or the respiration signal was so weak
that the monitor cannot perform respiration analysis. Check the
patient’s condition and the RM connections.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...