16-3
16.4
Preparation for CCO Monitoring
Please refer to the following figure and procedure to set up the CCO monitoring:
1.
Arterial thermodilution catheter
2.
Blood temperature sensor
3. Central
venous
catheter
4. PiCCO
cable
5.
Injectate temperature sensor cable
6.
Injectate temperature sensor
7. IBP
cable
8.
Arterial pressure transducer
9. CVP
transducer
1.
Place the arterial thermodilution catheter.
WARNING
The arterial thermodilution catheter must be placed in one of the bigger systemic arteries, for
example, the femoral, the brachial or the axillary artery.
You must use the approved catheters and puncture locations.
2.
Place the central venous catheter.
3.
Connect the injectate temperature sensor to the central venous catheter.
4.
Plug the PiCCO cable into the CCO/C.O. connector on the PiCCO module, and connect the following devices to the
PiCCO cable:
Injectate temperature sensor probe
Blood temperature sensor connector.
5.
Connect one end of the arterial pressure transducer to the arterial thermodilution catheter and the other end to the
IBP cable marked with pArt.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...