8-23
8.11
Resting 12-lead ECG Analysis
8.11.1
Entering the 12-lead Screen
1.
In the [
ECG Setup
] menu, select [
Others>>
] to enter the [
Other Setup Menu
].
2.
Set [
Lead Set
] to [
12-Lead
].
3.
Set [
ECG Display
] to [
12-Lead
].
The functions of the keys at the bottom of the 12-lead screen are as follows:
[
Analyze
]: starts resting 12-lead analysis.
[
Patient Info
]: enters the patient information.
[
Setup
]: enters the 12-lead setup menu.
[
Review
]: enters the [
Review
] window.
[
Exit
]: exits the 12-lead screen.
8.11.2
Entering Patient Information
Some patient information may directly affect ECG analysis. Complete and correct patient information is helpful for
accurate diagnosis and treatment of the patient. Enter patient information before taking an ECG measurement.
To enter the patient information, select [
Patient Info
] from the 12-lead screen.
NOTE
Check that patient information is correct before resting 12-lead analysis.
We recommend using pediatric lead placement V4R, V1, V2, V4 - V6 if the patient is under 16 years of age.
Please record V4R using the V3 electrode. Also set [V3 Electrode Placement] to [V4R]. This is a normal
practice for a patient of this age.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...