A-7
Common mode rejection ratio
(with Notch off )
Diagnostic mode:
Monitor mode:
Surgical mode:
ST mode:
>90 dB
>105 dB
>105 dB
>105 dB
Notch 50/60
Hz
Differential input impedance
≥5 MΩ
Input signal range
±8 mV (peak-to-peak value)
Accuracy of reappearing input signal
Use A and D methods based on EC11 to determine system total error and
frequency response.
Electrode offset potential tolerance
±500 mV
Lead-off detection current
Measuring electrode: <0.1 μA
Drive electrode: <1 μA
Input offset current
Measuring electrode: ≤0.1 μA
Drive electrode: ≤1 μA
Baseline recovery time
Enduring 5000V (360 J) charge without data loss or corruption
Baseline recovery time: <5 s (after defibrillation)
Polarization recovery time: <10 s
Defibrillation energy absorption: <10% (100Ω load)
Patient leakage current
<10 μA
Calibration signal
1mV (peak-to-peak value)
Accuracy:
±
5
%
ESU protection
Cut mode: 300 W
Coagulate mode: 100 W
Recovery time: ≤10 s
In compliance with the requirements in clause 202.6.2.101 of IEC60601-2-27
Pace Pulse
Pace pulse markers
Pace pulses meeting the following conditions are labelled with a PACE marker:
Amplitude:
Width:
Rise time:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 μs
Pace pulse rejection
When tested in accordance with theIEC60601-2-27: 201.12.1.101.13, the heart rate
meter rejects all pulses meeting the following conditions.
Amplitude:
Width:
Rise time:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 μs
Sampling rate
500 samples/s (A/D)
500 samples/s (ECG algorithm)
Accuracy 2.44
μV/LSB
HR
Measurement range
Neonate:
Pediatric:
Adult:
15 to 350 bpm
15 to 350 bpm
15 to 300 bpm
Resolution 1
bpm
Accuracy
±1 bpm or ±1%, whichever is greater.
Sensitivity
200μV (lead II)
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...