20-2
20.3.2
Transferring Data by Different Means
NOTE
Never enter the data transfer mode when the patient monitor is in normal operation or performs
monitoring. You must re-start the patient monitor to exit the data transfer mode.
Transfer data via a crossover network cable
Before transferring data using a crossover network cable, do as follows:
1.
Connect the T1 to the T1 docking station.
2.
Connect one end of the crossover network cable to the T1 docking station and the other end to the PC.
3.
Set the IP address of the PC. This IP address must be in the same network segment with that of the patient monitor.
4.
Make sure that the data export system is active on the PC.
Then, follow this procedure to transfer data:
1. Select
[
Main Menu
]
→
[
Patient Data >>
]
→
[
Transfer Data
].
2. Select
[
Yes
] from the popup message box.
3.
Input the IP address already set on the PC.
4. Select
[
Start
] to start transferring data.
Transfer data within a LAN
Before transferring data within a LAN, do as follows:
1.
Connect the patient monitor and the intended PC into the same LAN and acquire the PC’s IP address.
2.
Make sure that the data export system is active on the PC.
Follow the same procedure as via a crossover network cable to transfer data.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...