5.9.2
Transferring Patient via T1
T1 can be used with a host monitor or the docking station to implement patient transfer. For patient transfer via the
host monitor, refer to the host monitor’s operating manual for detail.
NOTE
Only host monitors with a system software version 05.00.00 or greater support patient transfer via T1.
In the situation that T1 is in use with the docking station for patient transfer, you need to configure the docking
station on T1. To configure the docking station, select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
[
Others >>
]
→
[
Dock Setup >>
].
NOTE
The function of dock setup is active only when T1 is inserted into the docking station.
Setting Woke Mode
Work Mode defines the source of department, bed number, and central station IP settings when T1 is connected with
the docking station.
In Dock mode, T1 uses the settings of T1 when T1 is connected, and the settings remain the unchanged when
T1 is removed from the docking station.
In host mode, T1 uses the settings of the docking station when T1 is connected, and the settings remains
unchanged when T1 is removed from the docking station.
Setting Network Type
Network Type specifies the source of T1’s network configuration when T1 is connected with the docking station.
If [
Use Current T1 Net Setting
] is selected, T1’s network setting is the current T1’s network setting when T1 is
connected, and the settings does not change when T1 is removed from the docking station.
If [
Use Current Dock Net Setting
] is selected, T1’s network setting is the current network setting of the
docking station, and the setting restore to the previous T1’s settings when T1 is removed from the docking
station.
Setting Central Station IP
Central Station IP defines the IP address of the central station to which T1 is connected. T1 can only be connected to
the specified central station.
The current central station IP is shown in the prompt message area. You can select this area to pop up the [
Central
Station IP
] menu.
NOTE
For central station with system software system software 06.08.00 or greater, as long as a patient is
admitted, T1 can be automatically connected to the specified central station.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...