A.2
Technical Alarm Messages
Measurement
Alarm message
L
I
Cause and solution
XX
XX SelfTest Err
H
C
An error occurred to the XX module, or there is a
problem with the communications between the
module and the monitor. Re-plug the module and
restart the monitor, or plug the module into another
monitor.
XX Init Err
H
A
XX Init Err N
H
A
N is within 1 to 8
XX Comm Err
H
A
XX Comm Stop
H
C
XX Comm Abnormal
H
A
XX Limit Err
L
C
XX parameter limit is accidentally changed. Contact
your service personnel.
XX Overrange
L
C
The measured XX value is not within the specified
range for XX measurement. Contact your service
personnel.
ECG
ECG Lead Off
L*
B
The electrode has become detached from the patient
or the lead wire has become disconnected from the
adapter cable. Check the connections of the electrodes
and leadwires.
ECG YY Lead Off
L*
B
Note: YY represents the leadwires V, LL, LA,
RA, C, F, L, R, V1, V2, V3, V4, V5, V6, C1, C2, C3,
C4, C5, or C6
ECG Noisy
L
A
The ECG signal is noisy. Check for any possible sources
of signal noise around the cable and electrode, and
check the patient for great motion.
ECG Artifact
(for Mortara algorithm only)
L
A
Artifacts are detected on the ECG analysis lead and as a
result heart rate cannot be calculated and Asystole,
Missed Beats and Vfib cannot be analyzed. Check the
connections of the electrodes and leadwires and check
for any possible source of interference around the
cable and electrode. Check the patient’s condition and
check the patient for great motion.
ECG Low Freq. Noise
L
A
Low frequency signals are detected on the ECG analysis
lead. Check for any possible source of interference
around the cable and electrode.
ECG Amplitude Too Small
L
C
The ECG amplitude didn’t reach the detected
threshold. Check for any possible source of interference
around the cable and electrode.
ECG Config. Err
L
C
ECG configuration is wrongly downloaded. Check the
downloaded configuration and re-download the
correct configuration.
Temp
Temp Cal. Err
H
C
A calibration failed. Restart the monitor.
YY Sensor Off
L
A
The temperature sensor has become detached from
the patient or the module. Check the sensor
connections.
YY represents a temperature label.
Temp Cal. Err
H
C
SpO
2
SpO
2
Sensor Off
L*
B
The SpO
2
sensor has become detached from the
patient or the module, or there is a fault with the SpO
2
SpO
2
Sensor Fault
L
C
D-3
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...