12-7
12.7.5
Switching On NIBP End Tone
The monitor can issue a reminder tone at the completion of NIBP measurement. The NIBP End Tone is off by default. You
can switch it on by accessing the [
NIBP Setup
] menu.
12.8
Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood vessel and therefore help venous
puncture.
1. Select
[
VeniPuncture >>
] from the [
NIBP Setup
] menu. In the popup menu, verify that the [
Cuff Press.
] value is
appropriate. Change it if necessary.
2. Select
[
VeniPuncture
].
3.
Puncture vein and draw blood sample.
4.
Select the on-screen QuickKey
to deflate the cuff. The cuff deflates automatically after a set time if you do not
deflate it.
During measurement, the NIBP display shows the inflation pressure of the cuff and the remaining time in venous
puncture mode.