8-17
8.8
About Arrhythmia Monitoring
Arrhythmia analysis provides information about your patient’s condition, including heart rate, PVC rate, rhythm and
ectopics.
WARNING
Arrhythmia analysis program is intended to detect ventricular arrhymias and atrial fibrillation. It is not
designed to detect all the atrial or supraventricular arrhythmias. It may incorrectly identify the presence or
absence of an arrhythmia. Therefore, a physician must analyze the arrhythmia information with other
clinical findings.
Heart-rate reading may be affected by cardiac arrhythmias. Do not rely entirely on heart rate alarms when
monitoring patients with arrhythmia. Always keep these patients under close surveillance.
Atrial fibrillation (Afib) detection function is not intended for pediatric and neonatal patients.
Mortara arrhythmia algorithm is not intended for neonatal patients.
8.8.1
Understanding the Arrhythmia Events
Mindray algorithm
Arrhythmia message
Description
Category
Asystole
No QRS detected within the set time threshold in absence of ventricular
fibrillation or chaotic signal.
Lethal
arrhythmia
Vfib/Vtac
A fibrillatory wave for 6 consecutive seconds.
A dominant rhythm of adjacent Vs and a HR > the V-Tac HR limit.
Vtac
The consecutive PVCs ≥ Vtac PVCs limit, and the HR ≥ the Vtac rate limit.
Vent. Brady
The consecutive PVCs ≥ the Vbrd threshold and the ventricular HR < the
Vbrd Rate threshold.
Extreme Tachy
The heart rate is equal to or greater than the extreme tachycardia limit.
Extreme Brady
The heart rate is equal to or less than the extreme bradycardia limit.
PVCs
PVCs/min exceeds high limit
Nonlethal
arrhythmia
Pacer not paced
No pace pulse detected for 1.75 x average R-to-R intervals following a
QRS complex (for paced patients only).
Pacer not capture
No QRS complex detected for 300 milliseconds following a pace pulse
(for paced patients only).
PVC
One PVC detected in normal heartbeats.
Couplet
Paired PVCs detected in normal heartbeats.
Run PVCs
More than 2 consecutive PVCs.
Bigeminy
A dominant rhythm of N, V, N, V, N, V.
Trigeminy
A dominant rhythm of N, N, V,N, N, V, N, N, V.
R on T
R on T detected in normal heartbeats.
Missed Beats
No beat detected for 1.75 x average R-R interval for HR <120, or
No beat for 1 second with HR > 120 (for non-paced patients only), or
No beat detected for more than the set pause threshold.
Brady
The average heart rate is equal to or less than the bradycardia limit.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...