A-3
A.3
Power Supply Specifications
External DC power supply
Input voltage
12 VDC (±10%)
Input current
2 A
DC adapter
Input
100 to 240 VAC (±10%), 50/60 Hz (±3 Hz)
Output
12VDC, minimum 2A
T1 docking station
Input voltage
100 to 240VAC (±10%)
Input current
0.65A to 0.35A
Frequency 50/60Hz
(±3Hz)
Battery
Battery Type
Chargeable Lithium-Ion
Capacity 2500
mAh
Charge time
Charge battery with the patient monitor:
Less than 3 hours to 90% and less than 4 hours to 100% with equipment power off;
Less than 12 hours to 90% and less than 14 hours to 100% with equipment power
on.
Charge battery with the battery charger:
Less than 3 hours to 90%
Less than 4 hours to 100%
Run time
when powered by a new fully-charged battery at 25
℃
with ECG and SpO
2
cable
connected, and auto NIBP measurements at an interval of 15 minutes:
5 h if charged by a battery charger
4.5 h if charged by installing in the monitor
Battery run time varies as per system configuration and settings.
Shutdown delay
at least 20 minutes (after a low battery alarm first occurs)
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...