C-1
C
Default Configurations
This chapter lists some of the most important factory default settings in configuration management. You cannot change
the factory default configuration itself. However, you can make changes to the settings from the factory default
configuration and then save the changed configuration as a user configuration.
C.1
Parameters Configuration
C.1.1
ECG
ECG Setup
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
Lead Set
Yes
Yes
Auto (automatic lead type identification)
Alm Source
Yes
Yes
HR
Alarm Yes
Yes On
Alm Lev
Yes
Yes
Med
HR/PR
High
Adu
Yes Yes
120
Ped 160
Neo 200
HR/PR
Low
Adu
Yes Yes
50
Ped 75
Neo 100
Sweep Yes
Yes 25
mm/s
Beat Vol
Yes
Yes
General, OR: 2
ICU, NICU, CCU: 1
Paced No
Yes No
Notch Filter
Yes
Yes
Weak
Gain Yes
Yes X1
Filter Yes
Yes
General: Monitor
OR: Surgery
ICU, NICU: Monitor
CCU: Diagnostic
ECG Display
Yes
Yes
Normal
Pacemaker Rate
Yes
Yes
60
Minimum QRS Threshold
No
Yes
0.16 mV
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...