D-6
Measurement
Alarm message
L
I
Cause and solution
CO
2
FilterLine Occluded
L
C
The airway or watertrap was occluded. Check the
airway and remove the occlusion.
CO
2
No Watertrap
L
B
Check the watertrap connections.
CO
2
Check Adapter
L
A
There is a problem with the airway adapter. Check,
clean or replace the adapter.
CO
2
FilterLine Err
L
C
Check if there is a leak in the CO
2
sample line or the CO
2
sample line has been occluded.
CO
2
Zero Failed
L
A
Check the CO
2
connections. After the sensor’s
temperature becomes stabilized, perform a zero
calibration again.
CO
2
System Err
L
A
Re-plug the module or restart the monitor.
CO
2
Check Cal.
L
C
Perform a calibration.
CO
2
Check Airway
L
C
An error occurred to the airway.
CO
2
No Filterline
L
A
Make sure that the filterline is connected.
CO
2
No Sensor
L
A
Make sure that the sensor is connected.
CO
2
Main Board Err
H
C
There is a problem with the CO
2
module. Re-plug the
module or restart the monitor.
CO
2
Checking Sensor
L
C
CO
2
Replace
Scrubber&Pump
L C
CO
2
15V Overrange
H
C
CO
2
Hardware Err
H
C
Power
12V Too High
H
C
There is a problem with the system power supply.
Restart the monitor.
12V Too Low
H
C
5V Too High
H
C
5V Too Low
H
C
3.3V Too High
H
C
3.3V Too Low
H
C
No Battery
H
C
1. Battery is not installed or poor contact: properly
install the battery.
2. Battery circuit failure or battery failed: contact your
service personnel.
Low Battery
M
C
Connect the monitor to external power source, or
charge the battery using a battery charger.
Battery Depleted
H
C
Connect the monitor to external power source, or
charge the battery using a battery charger.
T1 battery to be protected
and not work.
H
C
T1 battery will be soon protected and will not supply
power. If you are going to use T1 for patient transport,
please replace the battery.
T1 battery aged. Replace the
battery
L
C
T1 battery lifetime is expired. Replace the battery with
a new one.
Power Board Comm Err
H
C
Restart the monitor. If the problem persists, contact
your service personnel.
RT Clock Need Reset
L
C
Internal backup battery cell fails. Contact your service
personnel.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...