8-20
Mortara algorithm
Arrh. event
Range
Default
Step
Unit
PVCs High
1 to 100
10
1
/min
Asys. Delay
2 to 10
5
1
s
Vtac Rate
100 to 200
130
5
bpm
Vtac PVC
3 to 12
6
1
beats
Multif. PVC
3 to 31
15
1
beats
Tachy High
Adult: 100 to 300
Pediatric: 160 to 300
Adult: 100
Pediatric: 160
5 bpm
Brady Low
Adult: 15 to 60
Pediatric: 15 to 80
Adult: 60
Pediatric: 80
5 bpm
8.8.4
Setting the Extended Arrh. (For Mindray Algorithm Only)
The following arrhythmia events are defined as extended arrhythmia:
Extreme Tachy
Extreme Brady
Vent. Brady
Nonsus. Vtac
Multif. PVC
Irr. Rhythm
Pause
Afib
You can select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the required password
→
select
[
Alarm Setup >>
], and set [
Extended Arrh.
] to [
Enable
] or [
Disable
]. When [
Extended Arrh.
] is set to [
Disable
], the
patient monitor does not analysis the extended arrhythmia events and corresponding alarms are not given.
8.8.5
Reviewing Arrhythmia Events
Please refer to the
Review
chapter.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...