C-8
PA Alarm Limit
PA-S High (mmHg)
Yes
Yes
Adu:
Ped, Neo:
35
60
PA-S Low (mmHg)
Yes
Yes
Adu:
Ped, Neo:
10
24
PA-M High (mmHg)
Yes
Yes
Adu:
Ped, Neo:
20
26
PA-M Low (mmHg)
Yes
Yes
Adu:
Ped, Neo:
0
12
PA-D High (mmHg)
Yes
Yes
Adu:
Ped, Neo:
16
4
PA-D Low (mmHg)
Yes
Yes
Adu:
Ped, Neo:
0
-4
CVP, LAP, RAP, ICP, UVP, P3-P4 Venous Pressure Alarm Limits
IBP-M High (mmHg)
Yes
Yes
Adu:
Ped, Neo:
10
4
IBP-M Low (mmHg)
Yes
Yes
0
CPP Alarm Limits
CPP High (mmHg)
Yes
Yes
Adu:
Ped:
Neo:
130
100
90
CPP Low (mmHg)
Yes
Yes
Adu:
Ped:
Neo:
50
40
30
Art, Ao, BAP, FAP, LV, P1-P2 Arterial Pressure Scale
Scale (mmHg)
Yes
Yes
0-160
PA Scale
Scale (mmHg)
Yes
Yes
0-30
CVP, LAP, RAP, ICP, UVP Scale
Scale (mmHg)
Yes
Yes
0-20
UAP, P3-P4 Venous Pressure Scale
Scale (mmHg)
Yes
Yes
0-80
IBP Overlapping Left Scale
Scale (mmHg)
Yes
Yes
0-160
IBP Overlapping Right Scale
Scale (mmHg)
Yes
Yes
0-20
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...