E-1
E
Electrical Safety Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive maintenance program.
They are a proven means of detecting abnormalities that, if undetected, could prove dangerous to either the patient or
the operator. Additional tests may be required according to local regulations.
All tests can be performed by using commercially available safety analyzer test equipment. These procedures assume
the use of a 601PROXL International Safety Analyzer or equivalent safety analyzer. Other popular testers complying with
IEC 60601-1 used in Europe, such as Fluke, Metron, or Gerb, may require modifications to the procedure. Please follow
the instructions of the analyzer manufacturer.
The electrical safety inspection should be periodically performed every two years .The safety analyzer also proves to be
an excellent troubleshooting tool to detect abnormalities of line voltage and grounding, as well as total current loads.
E.1
Power Cord Plug
Test Item
Acceptance Criteria
The power plug
The power plug pins
No broken or bent pin. No discolored pins.
The plug body
No physical damage to the plug body.
The strain relief
No physical damage to the strain relief. No plug
warmth for device in use.
The power plug
No loose connections.
The power cord
No physical damage to the cord. No deterioration to
the cord.
For devices with detachable power cords, inspect the
connection at the device.
For devices with non-detachable power cords, inspect
the strain relief at the device.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...