A-12
A.8.4
PR
Alarm limit
Range (bpm)
Step (bpm)
PR High
(low limit +2) to 300
1
PR Low
15 to (high limit-2)
PR from Mindray SpO
2
Module
Measurement range
20 to 300 bpm
Resolution 1
bpm
Response time
≤ 30 s (PI > 0.3, no disturbance, PR value sudden change within 25 – 250 bpm)
Accuracy ±3
bpm
Refreshing rate
1 s
PR from Masimo SpO
2
Module
Measurement range
25 to 240 bpm
Resolution 1
bpm
Response time
≤ 30 s (PI > 0.3, no disturbance, PR value sudden change within 25 – 240 bpm)
Accuracy
±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refreshing rate
≤ 2 s
Low perfusion conditions
Pulse amplitude: >0.02%
Light penetration: >5%
Low perfusion PR accuracy
±3 bpm
PR from Nellcor SpO
2
Module
Measurement range
20 to 300 bpm
Resolution 1
bpm
Response time
≤ 30 s (PI > 0.3, no disturbance, PR value sudden change within 25 – 250 bpm)
Accuracy
20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refreshing rate
≤ 2 s
PR from IBP Module
Measurement range
25 to 350 bpm
Resolution 1
bpm
Accuracy
±1 bpm or ±1%, whichever is greater
Refreshing rate
1 s
PR from NIBP Module
Measurement range
30 to 300 bpm
Resolution 1
bpm
Accuracy
±3bpm or ±3%, whichever is greater
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...