ELWI < 10
ELWI > 10
ELWI < 10
Patient Weight (kg)
Iced Injectate
Iced Injectate
Room Temperature Injectate
<3 2ml
2ml
3ml
<10 2ml
3ml
3ml
<25 3ml
5ml
5ml
<50 5ml
10ml
10ml
<100 10ml
15ml
15ml
≥
100
15ml 20ml
20ml
11. Set up the C.O. measure mode by selecting [
C.O. Measure
] from the
[
CCO Setup
] menu, and toggling between
[
Auto
]
and [
Manual
].
If you select [
Manual
], you should start each measurement manually by pressing the [
Start
] key in the [
PiCCO
Measurement
].
If you select [
Auto
], the C.O. measurements can be performed consecutively, without the need for pressing the
[
Start
] key.
16.5
Performing CCO Monitoring and CCO Calibration
Please perform CCO monitoring according to the following procedure:
1.
Enter the [
PiCCO Measurement
] window.
A. Thermodilution
curve
B.
Prompt message area
C. Buttons
D. History
window
E.
Measurement quality:
△
T
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...