14-12
14.8.4
Performing Hemodynamic Calculation
In the PAWP window, select [Calc.>>] to enter the hemodynamic calculation menu. Refer to
18.5 Hemodynamic
Calculations
for details.
14.9
Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or
accessories, check the table below before requesting for services. If the problem persists, contact your service personnel.
CAUTION
Never try to disassemble the equipment or supplied accessories. There are no internal user-serviceable
parts.
Symptoms
Possible Cause
Correction Action
Damped invasive
waveform
Air bubbles in tubing
Eliminate air from tubing as described in
14.4
Setting Up the Pressure Measurement.
Kinked catheter
Change the position of catheter.
Blood in tubing
Pressurize the solution bag to 300 mmHg. For
details, refer to the instructions for use of the
solution bag.
IBP not displayed or no IBP
waveform
Improper setup
Check display setup in monitor setup.
Cable not plugged in
Check that the cables are properly connected.
Transducer not connected.
Check that the transducer is properly
connected.
Stopcock turned improperly.
Check that the stopcock is turned to the correct
position.
Transducer not zeroed
Check and zero the transducer as described in
14.3 Zeroing the Transducer
.
Dashes “- -” display in place
of numerics.
The measured result is invalid or out of
range.
IBP might be set to non-pulsatile labels like
CVP, LA, RA, and ICP.
Change to a pulsatile label.
Abnormally high or low
readings
Transducer too High or too Low.
Adjust the position of the transducer and make
sure that it is level with the heart,
approximately at the level of the midaxillary
line.
Zero the transducer as described in
14.3
Zeroing the Transducer
Unable to Zero
Stopcock not open to atmosphere.
Check the transducer and make sure the
stopcock is turned to the air.
PAWP button disabled
One IBP channel must be labeled PA
Label an IBP channel as PA. (Also Label an IBP
channel as P1/P2/P3/P4, it will automatically
change to PA)
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...