Setting Pacemaker Rate (For Mortara only)
Some pacemaker pulses can be difficult to reject. When this happens, the pulses are counted as a QRS complex and
could result in an incorrect HR and failure to detect some arrhythmias. You can set [
Pacemaker Rate
] to the pacemaker’s
rate in the [
ECG Setup
] menu. In this way, the patient monitor can calculate HR and detect arrhythmias more accurately.
When [
Paced
] is set to [
No
], the pacemaker rate cannot be set.
8
.5.6
Choosing an ECG Display Screen
When monitoring with a 5-lead or 12-lead set, you can select [
Others>>
]
→
[
ECG Display
]in the [
ECG Setup
] menu to
choose the screen type as:
[
Normal Screen
]: The ECG waveform area shows 2 ECG waveforms.
[
Full-Screen
]: The whole waveform area shows 7 ECG waveforms only.
When monitoring with a 12-lead set, you can also choose the screen type as [
12-Lead
].
When the screen type is set to [
Normal Screen
], cascaded ECG waveforms can be displayed. To cascade ECG waveforms:
1.
Select the [
Main Menu
]
→
[
Screens
]
→
[
Screen Setup
].
2. Select
[
ECG1 Casc.
] in the second row. A cascaded waveform is displayed in two waveform positions.
8
.5.7
Setting the Notch Filter
The notch filter removes the line frequency interference. Only when [
Filter
] is set to [
Diagnostic
], the [
Notch Filter
] is
adjustable.
1.
Select the ECG parameter window or waveform area to enter its setup menu. Then select [
Others >>
].
2. Set
[
Notch Filter
] to
[
Strong
] when there is strong interference (such as spikes) with the waveform.
[
Weak
] when there is weak interference with the waveform.
[
Off
] to turn the notch filter off.
Set notch frequency according to the electric power frequency of your country. Follow this procedure:
1. When
[
Notch Filter
] is set on, select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the
required password.
2. Select
[
Others >>
]
→
[
Notch Freq.
] and then select [
50Hz
] or [
60Hz
] according to the power line frequency.
8
.5.8
Changing the Pacer Reject Settings
Select [
ECG Setup
]
→
[
Others>>
]
→
[
Pacer Reject
], and toggle between [
On
] and [
Off
].
When [
Pacer Reject
] is switched on, pace pulses are not displayed..
When [
Pacer Reject
] is switched off, pace pulses are displayed.
NOTE
When pace pulses are detected, pace pulse marks “|” are shown on the ECG waveforms. Pacer Rejection
setting has no impact on the display of pace pulse marks “|”
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...