6.10
Restoring the Latest Configuration Automatically
During operation, you may make changes to some settings. However, these changes may not be saved as user
configuration. To prevent the changes from losing in case of a sudden power failure, the patient monitor stores the
configuration in real time. The saved configuration is the latest configuration.
The monitor restore the latest configuration if restarts within 60 seconds after the power failure. And it will restore the
default configuration rather than the latest configuration if restarts 120 seconds later after the power failure. The
monitor may load either the latest configuration or the default configuration if restarts from 60-120 seconds after the
power failure.
6.11
Modifying Password
To modify the password for accessing the [
Manage Configuration
] menu,
1. Select
[
Modify Password >>
] in the [
Manage Configuration
] menu.
2.
Input a new password in the popup menu.
3. Select
[
Ok
].
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...