23-8
WARNING
Connect an exhaust tube to the gas outlet connector of the monitor to vent the calibration gases to a
scavenging system.
23.10
Calibrating the Touchscreen
1. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
Cal. Touchscreen
]. The
symbol will appear at different positions
of the screen.
2.
Select, in turn the central point of the
symbol. After the calibration is completed, the message [
Screen
Calibration Completed!
] is displayed.
3. Select
[
Ok
] to confirm the completion of the calibration.
You cannot calibrate the touchscreen if it is locked.
23.11
Electrical Safety Tests
Refer to
E Electrical Safety Test
.