C-3
Arrhythmia Threshold Settings (Mortara)
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
PVCs high
Yes
Yes
10
Asys. Delay
Yes
Yes
5
Vtac Rate
Yes
Yes
130
Vtac PVCs
Yes
Yes
6
Multif. PVC’s Window
Yes
Yes
15
Tachy Yes
Yes
Adu:
Ped:
120
160
Brady Yes
Yes
Adu:
Ped:
50
75
Arrhythmia Alarm Settings (Mindray)
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
Alarm Switch
Alarm Level
Asystole Alarm
Yes
Yes
On
High
VFib/VTac Alarm
Yes
Yes
On
High
Vtac Alarm
Yes
Yes
On
High
Vent. Brady Alarm
Yes
Yes
On
High
Extreme Tachy Alarm
Yes
Yes
On
High
Extreme Brady Alarm
Yes
Yes
On
High
PVCs/min Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
R on T Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Run PVCs Alarm
Yes
Yes
Off
Low
Couplet Alarm
Yes
Yes
Off
Prompt
Multif. PVC Alarm
Yes
Yes
Off
Med
PVC Alarm
Yes
Yes
Off
Prompt
Bigeminy Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Trigeminy Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Tachy Alarm
Yes
Yes
Off
Med
Brady Alarm
Yes
Yes
Off
Med
PNP Alarm
Yes
Yes
Off
Prompt
PNC Alarm
Yes
Yes
Off
Prompt
Missed Beats Alarm
Yes
Yes
Off
Prompt
Nonsus. Vtac Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Vent. Rhythm Alarm
Yes
Yes
Med
Pause Alarm
Yes
Yes
Off
Low
Irr. Rhythm Alarm
Yes
Yes
Off
Prompt
Afib Alarm
Yes
Yes
Off
Prompt
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...