Symptoms
Possible Cause
Correction Action
Changing ECG Wave Settings
.
Lead wires and patient cable not fully
or properly inserted
Check that the leadwires and patient cables are
properly connected.
Cable or lead wires damaged
Change cable and lead wires.
Base Line Wander
Patient moving excessively
Secure leadwires and cable to patient.
Electrodes dry or loose
Repeat skin preparation as described in
8.3.1
Preparing the Patient and Placing the Electrodes
and apply fresh and moist electrodes.
ECG Filter set to ST or Diagnostic
mode
Set ECG Filter to “Monitor” mode.