15-7
15.4.7
Setting the Flow Rate
For the sidestream CO
2
module, you can change the sampling rate of respiratory gas in the patient’s airway by setting
the flow rate. To set the flow rate, enter the [
CO2 Setup
] menu and select an appropriate setting from [
Flow Rate
].
WARNING
Please consider the patient’s actual bearing capability and select the appropriate flow rate when setting the
flow rate.
15.4.8
Setting up the CO
2
Wave
In the [
CO2 Setup
] menu, you can:
Select [
Wave Type
] and toggle between [
Draw
] and [
Fill
]:
[
Draw
]: The CO
2
wave is displayed as a curved line.
[
Fill
]: The CO
2
wave is displayed as a filled area.
Select [
Sweep
] and then select the appropriate setting. The faster the wave sweeps, the wider the wave is.
Change the size of the CO
2
waveform by adjusting the wave [
Scale
].
15.4.9
Setting RR Source
To set RR source:
1.
Enter the [
CO2 Setup
] menu.
2. Select
[
RR Source
] and then select a source or [
Auto
] from the dropdown list.
The [
RR Source
] settings of Resp and CO
2
, are linked. For details, please refer to
Setting RR Source of chapter Resp
.
15.4.10
Setting Barometric Pressure Compensation
Both sidestream and microstream CO
2
modules have the function of automatic barometric pressure compensation (the
system automatically measures the barometric pressure which the patient monitor is exposed to). However, the
mainstream CO
2
module does not have such function. For the mainstream CO
2
module, the default barometric pressure
is 760 mmHg. You must modify the barometric pressure based on the actual situation as follows:
1. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the required password
→
[
Module
Maintenance >>
]
→
[
Maintain CO2 >>
]
→
[
Calibrate CO2 >>
].
2. Select
[
Barometric Pressure
] and then enter the value of barometric pressure to which the patient monitor is
exposed to.
WARNING
Be sure to set the barometric pressure properly before using the mainstream CO
2
module. Improper settings
will result in erroneous CO
2
reading.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...