B-1
B
EMC and Radio Regulatory Compliance
B.1
EMC
The device meets the requirements of IEC 60601-1-2: 2014.
WARNING
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this device could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this device and result in improper operation.
Use of this device adjacent to or stacked with other device should be avoided because it could
result in improper operation. If such use is necessary, this device and the other device should
be observed to verify that they are operating normally.
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the this
device, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this device could result.
The non-ME EQUIPMENT (e.g. ITE) that is a part of an ME SYSTEM may be disrupted by the
electromagnetic interference of nearby equipment. It may be necessary to take mitigation
measures, such as re-orienting or relocating the non-ME EQUIPMENT or shielding the location.
This device is intended for use in professional healthcare facility
EMC
environment and
home
healthcare
EMC
environment
only
. If it is used in special environment, such as magnetic
resonance imaging environment, the equipment/system may be disrupted by the operation of
nearby equipment.
Guidance and Declaration - Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user
of the device should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
Conducted and radiated RF
EMISSIONS
CISPR 11
Group 1
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic device.
Conducted and radiated RF
EMISSIONS
CISPR 11
Class B
(during
patient
transport)
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Conducted and radiated RF
Class A
The device is suitable for use in all establishments
other than domestic and those directly connected to
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...