7-15
WARNING
Resetting care group alarms may cause a potential hazard. Please act with caution.
7.10.3
Switching Off the Remote Device Disconnection Alarm
The monitor can provide an alarm if a viewed bed device is disconnected. By default, the function is enabled. To disable
the alarm, follow this procedure:
1.
In the current monitor, select [
Main Menu
]
→
[
Maintenance>>
]
→
[
User Maintenance>>
]
→
enter the required
password
→
[
Alarm Setup>>
]
→
[
Other Bed Alarm Setup>>
].
2.
Set [
Other Bed Disconnection Alm
] to [
Off
].
7.10.4
Setting Care Group Alert Tone
7.10.4.1
Setting the Alarm Reminder
When
a monitor in the Care Group issues an alarm, your patient monitor prompts you by giving alert tone. To set the
alert tone, follow this procedure:
1.
In the current monitor, select [
Main Menu
]
→
[
Maintenance>>
]
→
[
User Maintenance>>
]
→
enter the required
password
→
[
Alarm Setup>>
]
→
[
Other Bed Alarm Setup>>
].
2.
Set the [
Alarm Reminder
].
[
Repeat
]: The monitor gives continuous alert tone when the alarm occurs at the viewed bed is the same level
as the setup level in the monitor. To set which alarm level applies to continuous alert tone, see section
7.10.4.2Setting the Alarm Level
.
[
Once
]: The monitor gives a single alert tone when an alarm occurs at the viewed bed.
[
Off
]: The monitor do not give any alert tone when an alarm occurs at the viewed bed.
7.10.4.2
Setting the Alarm Level
When [
Alarm Reminder
] is set to [
Repeat
], you can set which alarm level of the viewed bed alarm applies to the
continuous alert tone. To set the alarm level of the viewed bed alarm, follow this procedure:
1.
In the current monitor, select [
Main Menu
]
→
[
Maintenance>>
]
→
[
User Maintenance>>
]
→
enter the required
password
→
[
Alarm Setup>>
]
→
[
Other Bed Alarm Setup>>
].
2.
Set the [
Alarm Lev
].
[
All
]: This monitor gives continous alert tone to all the alarms of the viewed bed when [
Alarm Reminder
] is set
to [
Repeat
].
[
High Only
]: This monitor gives continous alert tone only to high level alarms of the viewed bed when [
Alarm
Reminder
] is set to [
Repeat
].
[
High&Med
]: This monitor gives continous alert tone to high level and mediate level alarms of the viewed bed
when [
Alarm Reminder
] is set to [
Repeat
].
NOTE
The setting of the [Alarm Lev] is valid only when [Alarm Reminder] is set to [Repeat].
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...